Optimization of Phacoemulsification Tip Gauge on the Oertli CataRhex3 in an in vitro Setting
Optimization of Phacoemulsification Tip Gauge on the Oertli CataRhex3 in an in vitro Setting
About this item
Full title
Author / Creator
Publisher
New Zealand: Dove Medical Press Limited
Journal title
Language
English
Formats
Publication information
Publisher
New Zealand: Dove Medical Press Limited
Subjects
More information
Scope and Contents
Contents
To evaluate the effect of each of the tip sizes available for the Oertli CataRhex3
phacoemulsification machine on efficiency.
Porcine lenses were fixed in formalin for 2 hours, then cut into 3.0 mm cubes. We studied three Oertli tips, all of which had a 30-degree bevel: easyTip 2.2 mm (20G), easyTip 2.8 mm (19G), and CO-MICS (21G). For the 19...