Unheralded failure of Riata defibrillator lead identified at defibrillation threshold testing
Unheralded failure of Riata defibrillator lead identified at defibrillation threshold testing
About this item
Full title
Author / Creator
Publisher
Tokyo: Elsevier B.V
Journal title
Language
English
Formats
Publication information
Publisher
Tokyo: Elsevier B.V
Subjects
More information
Scope and Contents
Contents
Abstract Background St Jude Medical “Riata” implantable cardioverter-defibrillator leads have a higher-than-expected failure rate and carry a Class 1 recall. Failure usually relates to an insulation breach that may not be identified by lead fluoroscopic or electrical changes. Case presentation We report a case of Riata lead failure identified by th...