Accuracy of Linear Depolarisation Ratios in Clean Air Ranges Measured with POLIS-6 at 355 and 532 NM
Accuracy of Linear Depolarisation Ratios in Clean Air Ranges Measured with POLIS-6 at 355 and 532 NM
About this item
Full title
Author / Creator
Publisher
Les Ulis: EDP Sciences
Journal title
Language
English
Formats
Publication information
Publisher
Les Ulis: EDP Sciences
Subjects
More information
Scope and Contents
Contents
Linear depolarization ratios in clean air ranges were measured with POLIS-6 at 355 and 532 nm. The mean deviation from the theoretical values, including the rotational Raman lines within the filter bandwidths, amounts to 0.0005 at 355 nm and to 0.0012 at 532 nm. The mean uncertainty of the measured linear depolarization ratio of clean air is about...
Alternative Titles
Full title
Accuracy of Linear Depolarisation Ratios in Clean Air Ranges Measured with POLIS-6 at 355 and 532 NM
Authors, Artists and Contributors
Author / Creator
Identifiers
Primary Identifiers
Record Identifier
TN_cdi_doaj_primary_oai_doaj_org_article_3133f61b30364d4bac118f47cf3824b7
Permalink
https://devfeature-collection.sl.nsw.gov.au/record/TN_cdi_doaj_primary_oai_doaj_org_article_3133f61b30364d4bac118f47cf3824b7
Other Identifiers
ISSN
2100-014X,2101-6275
E-ISSN
2100-014X
DOI
10.1051/epjconf/201611925013