Limitation of the bandpass filter in preventing oversensing of pectoral myopotentials over the long‐...
Limitation of the bandpass filter in preventing oversensing of pectoral myopotentials over the long‐term follow‐up
About this item
Full title
Author / Creator
Publisher
Japan: John Wiley & Sons, Inc
Journal title
Language
English
Formats
Publication information
Publisher
Japan: John Wiley & Sons, Inc
Subjects
More information
Scope and Contents
Contents
A 60‐year‐old male experienced an inappropriate shock from an implantable cardioverter‐defibrillator (ICD) because of oversensing of pectoral myopotentials. Battery depletion was also observed, and a generator change was performed. A single‐chamber ICD (VENTAK PRIZM II 1860) was changed to a new ICD (INCEPTA VR F161). The myopotentials were clearly...