A patient with an implantable cardioverter‐defibrillator (ICD) (Abbott®) had episodes of slow monomorphic ventricular tachycardia (VT) and his ICD was programmed with three tachycardia zones. During the follow‐up, he received an inappropriate shock. Upon interrogation (of the device), trigeminal pattern binned as ventricular sensing (VS)—VS—ventric...