Log in to save to my catalogue

Do We Deliver the Pressures We Intend to When Using a T-Piece Resuscitator?

Do We Deliver the Pressures We Intend to When Using a T-Piece Resuscitator?

https://devfeature-collection.sl.nsw.gov.au/record/TN_cdi_plos_journals_1354337158

Do We Deliver the Pressures We Intend to When Using a T-Piece Resuscitator?

About this item

Full title

Do We Deliver the Pressures We Intend to When Using a T-Piece Resuscitator?

Publisher

United States: Public Library of Science

Journal title

PloS one, 2013-05, Vol.8 (5), p.e64706-e64706

Language

English

Formats

Publication information

Publisher

United States: Public Library of Science

More information

Scope and Contents

Contents

A T-piece resuscitator (TPR) uses a built-in manometer to set the inflation pressures, but we are not informed what pressures are actually delivered distally. Aim of this study was to measure the proximal and distal pressures under different gas conditions when using a TPR.
A test lung was ventilated using a TPR (PIP 25 cmH₂O, PEEP 5 cmH₂O) with...

Alternative Titles

Full title

Do We Deliver the Pressures We Intend to When Using a T-Piece Resuscitator?

Identifiers

Primary Identifiers

Record Identifier

TN_cdi_plos_journals_1354337158

Permalink

https://devfeature-collection.sl.nsw.gov.au/record/TN_cdi_plos_journals_1354337158

Other Identifiers

ISSN

1932-6203

E-ISSN

1932-6203

DOI

10.1371/journal.pone.0064706

How to access this item