Log in to save to my catalogue

Optimization Sonication Time and Dilution Factor in Determining the Concentration of Endotoxin Chall...

Optimization Sonication Time and Dilution Factor in Determining the Concentration of Endotoxin Chall...

https://devfeature-collection.sl.nsw.gov.au/record/TN_cdi_proquest_journals_2512976464

Optimization Sonication Time and Dilution Factor in Determining the Concentration of Endotoxin Challenge Vial with Kinetic Turbidimetric Method

About this item

Full title

Optimization Sonication Time and Dilution Factor in Determining the Concentration of Endotoxin Challenge Vial with Kinetic Turbidimetric Method

Author / Creator

Publisher

Bristol: IOP Publishing

Journal title

Journal of physics. Conference series, 2021-02, Vol.1764 (1), p.12017

Language

English

Formats

Publication information

Publisher

Bristol: IOP Publishing

More information

Scope and Contents

Contents

ECV has a specification certificate of analysis (CoA) worth 9616 EU/mL where the limit of acceptability recovery is in the range of 50 – 200 percent (4808 EU/mL - 19232 EU/mL). The verification results of ECV concentration with kinetic turbidimetry method showed recovery outside the lower limit of the predetermined range less than 50 percent. This...

Alternative Titles

Full title

Optimization Sonication Time and Dilution Factor in Determining the Concentration of Endotoxin Challenge Vial with Kinetic Turbidimetric Method

Authors, Artists and Contributors

Identifiers

Primary Identifiers

Record Identifier

TN_cdi_proquest_journals_2512976464

Permalink

https://devfeature-collection.sl.nsw.gov.au/record/TN_cdi_proquest_journals_2512976464

Other Identifiers

ISSN

1742-6588

E-ISSN

1742-6596

DOI

10.1088/1742-6596/1764/1/012017

How to access this item