Background:
 Coronary sinus (CS) lead placement for cardiac resynchronization therapy has a failure rate of ∼5–10%. Here we describe a way of implanting an endocardial left ventricular (LV) lead via a transseptal puncture (TSP), using a GooseNeck snare and active fixation lead.
Methods:
 Three male patients (67–83 years) with failed or ext...